Home | Careers | Contact Us | Product Search

(0) Items
View Cart
Check Out

Clothing & Accessories

Disinfection & Cleaning

Cleanroom Accessories

Monitoring

 

Data Loggers : VL-2000  

Validatable Relative Humidity Logger

  • Portable self-contained relative humidity & temperature data logger for critical validation applications
  • Equipped with internal relative humidity & temperature sensors, memory and ten-year battery
  • Tamper-proof operation produces reliable and secure documentation for validation records
  • Password-protected setup, download and calibration functions ensure integrity of collected data
  • Records relative humidity (RH) and temperature without wires, power cords or connections
  • Traceable instrument-grade RH and temperature measurement accuracy
  • Adjustable logging intervals with multi-year data recording capacity
  • Premium quality precision sensors offer exceptional in-calibration data logger performance

Technical Specifications

Internal Relative Humidity Sensor

Type of Sensor: Integrated high-accuracy capacitive polymer RH sensor
Initial Accuracy: +/-1%RH over 10%RH to 90%RH at +20ºC to +30ºC

+/-2%RH over 10%RH to 90%RH at -20ºC to +70ºC
Operating range: -35°C to 85°C (-31°F to 185°F) and 0% RH to 100% RH
Resolution: 0.05% RH
1-year Accuracy: +/-2% RH over 10% RH to 90% RH at +20°C to +30°C

+/-3% RH over 10% RH to 90% RH at -20°C to +70°C
Calibration verification: Annual (recommended)

Internal Temperature Sensor

Initial Accuracy: +/-0.10ºC over +20ºC to +30ºC

+/-0.15ºC over -25ºC to +70ºC
Operating range: +/-1%RH over 10%RH to 90%RH at +20ºC to +30ºC

+/-2%RH over 10%RH to 90%RH at -20ºC to +70ºC
Resolution: 0.02°C at 25°C
1-year Accuracy: +/-0.15°C over 20°C to +30°C

+/-0.25°C over -25°C to +70°C

General

Size: 2.8 x 2.1 x 0.8" (71x53x20mm); 65g (2.4 oz)
Operating Range: -35°C to 85°C (-31°F to 185°F) and 0% RH to 100% RH
Interfaces: RS-232 serial, USB, Ethernet, WiFi network interface available.
Mounting: Magnetic strips; 3M Dual Lock™ fasteners
PC Software: Designed for use with vLog Software and Veriteq viewLinc (for on-going monitoring & alarming)
Clock: Accuracy: +/- 1 min./month at 0 to 50°C
Electromagnetic Compatibility: FCC Part 15 and CE
Power Source: Internal lithium battery with life of 10 years at 1 minute sampling rate.

Memory

Memory Type : Non-volatile EEROM
Data Sample Capacity: 26,900 12-bit samples
Memory Modes: User-selectable: Wrap when memory full (FIFO), or Stop when memory full.
Memory Protection: Data retention > 20 years backup without power.
Sampling Rates: User-selectable (in intervals of 10 seconds) from once every 10 seconds to once a day.